Multi-Therapy Infusion Pump
Archived on April 29, 2026. Original response deadline: . Full description on SAM.gov.
Description
1.0 Product Overview: This document outlines the requirements for a small, lightweight, and durable multi-therapy infusion pump designed for use in pre-hospital, aeromedical, and military environments. The Office of the Command Surgeon is seeking a portable device that offers a wide range of therapeutic applications in a compact and rugged form factor, capable of reliable function in demanding field conditions. The intended use for this device is for intravascular, subcutaneous, intra-arterial, epidural, and perineural infusions administered by licensed military healthcare professionals in clinical, ambulatory, and medical transport settings (both air and ground).
2.0 Core Capabilities and Specifications: The primary requirements for the infusion pump are focused on size, weight, and cube (SWaP) to ensure it can be seamlessly integrated into already existing medical pack-outs and ruck sacks.
Feature Required Specification
- Dimensions Objective: Less than 6” x 4” x 2”
- Weight Less than 15 oz (excluding battery)
- Delivery Modes Must support: Continuous, Patient-Controlled Analgesia (PCA), Intermittent, Multi-Step, Total Parenteral (TPN), and Epidural infusions.
- Flow Rate 0.1 mL/h to 999 mL/h
- Flow Rate Accuracy ±2.5% under normal conditions
- Volume to be Infused 0.1 mL to 9,999 mL
- Drug Library Must support up to 40 clinical care areas and at least 1,000 drugs per area.
- Power Source AC: 100-240V, 50-60 Hz DC: Rechargeable Li-Ion battery
- Battery Life Minimum of 24 hours of continuous operation at a rate of 125 mL/h.
- Environmental Device must feature a touch screen that is fully operable when wet. The casing must be designed to withstand drops and fluid splashes.
- Operating Temperature +5°C to +40°C (41°F to 104°F)
- The device must meet the following minimum standards for military and aeromedical use:
- Electromagnetic Compatibility: Must meet MIL-STD-461G for Electromagnetic Interference (EMI) and Compatibility (EMC).
- Aeromedical Certification: Must be certified as "Safe-to-Fly" for use in military and commercial air transport.
- FDA Clearance: The device must have current FDA clearance for the indicated uses.
- Technical Considerations: The device must be capable of wired data connectivity. While a non-standard serial port is acceptable, it must be supplied with an appropriate adapter to ensure connectivity with modern computer systems (e.g., USB).
- Consumables: Please specify if the infusion pump system requires proprietary consumables (e.g., administration sets).
- Performance Testing: The device will be subject to a series of performance checks measuring the Volume to be Infused (VTBI) to validate accuracy and reliability.
- Submission: In response to this request for information, vendors shall submit a capability statement or white paper demonstrating that the proposed items meet all outlined specifications. Submissions must be received by the date specified in the SAM.gov announcement and submitted to the point of contact identified therein.
Overview
Award Overview
Possible prior contracts
Sign in to see possible prior contracts for this requirement (matched by agency, NAICS and place of performance; not confirmed).
Contacts
Contracting Officer
Primary Contact
Secondary Contact
Documents
No documents available.
Lifecycle
FA4417MTPUMP- Market Research
Pre-solicitation or formal solicitation typically follows within 30–180 days. (estimate)
- Market ResearchCurrentApr 8, 2026
Multi-Therapy Infusion Pump
Similar Opportunities
Title | Phase | Agency | NAICS | Set Aside | Posted | Deadline | Location |
|---|---|---|---|---|---|---|---|
DESCRIPTION: The VA Healthcare System (VAHCS) at Bay Pines, Florida, intends to award on a Sole Source basis. Seeks to procure Penumbra Lightning Flash intelligent aspiration catheters to support mechanical thrombectomy (clot removal) procedures performed by Bay Pines VAHCS providers. These single-use, FDA-cleared endovascular aspiration catheters are used to remove thrombus (blood clots) from vessels during emergent and scheduled vascular procedures, including the treatment of acute vascular occlusion. The catheters operate in conjunction with the facility's existing Penumbra System, including the ENGINE aspiration pump and associated aspiration tubing. This procurement will aid in the prompt, safe, and effective treatment of Veterans requiring thrombectomy. This is a one-time supply purchase. The proposed source is USVETSERV LLC., Greenville, SC. THIS NOTICE IS NOT A REQUEST FOR COMPETITIVE QUOTES; however, any firm that believes it can meet the requirements may give written notification prior to the response due date and time. The proposed NAICS 339112 is applicable to this acquisition; with a size standard of 1,000 employees. Parties may identify their interest, accompanied by supporting documentation to the Contracting Officer no later than 10/23/2026 at 4:30PM EST. Responses should be sent to yolanda.glover@va.gov. Responses received will be evaluated; however, a determination by the Government not to compete the proposed procurement will be solely up to the government. | Market Research | 339112 | Service-Disabled Veteran-Owned Small Business Set-Aside | 10/19/2026 | 10/23/2026 | - | |
The Orlando VA Healthcare System Audiology Department requires Drying Systems for Hearing Aids and Electronic Hearable Devices. 36C24826Q1194 0001 is hereby issued to answer questions that were received during the Q&A period. | Solicitation | 339112 | No Set-Aside | 09/02/2026 | 09/08/2026 | - | |
This is a presolicitation notice for Drying System for Hearing Aids and Electronic Hearable Devices for the Orlando VA Healthcare System. Please search for RFQ 36C24826Q1194 which will be posted shortly after this notice. | Pre-Solicitation | 339112 | Service-Disabled Veteran-Owned Small Business Set-Aside | 08/27/2026 | 09/08/2026 | - | |
The Orlando VA Healthcare System Audiology Department requires Drying Systems for Hearing Aids and Electronic Hearable Devices. | Solicitation | 339112 | No Set-Aside | 08/27/2026 | 09/08/2026 | - | |
GENERAL: James A. Haley Veterans Hospital (JAHVH) located at 13000 Bruce B Downs Blvd Tampa, Florida 33612, has a requirement for the procurement of the Philips Neximer Laser system for use within the Cardiology Section of Medicine and Specialty Care. PURPOSE: Neximer lasers are image-guided excimer laser systems used in interventional cardiology to treat a range of coronary and peripheral vascular problems, as well as certain cardiac lead management issues. SCOPE: JAHVH Tampa Medicine and Specialty Care is requesting Philips Neximer PHYSICAL CHARACTERISTICS: Equipment must have physical characteristics to enable the government to perform the capabilities in accordance with the Statement of Need RFQ Responses are due at 12:00PM on 8/6/2026. Questions in response to this solicitation may be emailed to Katrina.tavares@va.gov NO LATER THAN 2:00PM on 8/3/2026. | Combined Synopsis | 339112 | Service-Disabled Veteran-Owned Small Business Set-Aside | 07/29/2026 | 08/06/2026 | Tampa, FL |
Additional Details
Solicitation Number
FA4417MTPUMP
Archive Date
Apr 29, 2026
Office Address
HURLBURT FIELD, FL, 32544-5810