L&D Surgical Lights
Archived on July 7, 2026. Original response deadline: . Full description on SAM.gov.
Overview
- Response deadline
- 11:00 AM MDT
- Posted
- June 16, 2026
- Notice type
- Solicitation
- Set-aside
- SDVOSB Sole Source
- NAICS
- 339113: Surgical Appliance and Supplies Manufacturing
- Agency
- Department of WarDepartment of the Army
Notice details
- Office
- US ARMY MEDICAL COMMAND.US ARMY HEALTH CONTRACTING ACTIVITY.W40M MRC0 WEST
- Place of performance
- Colorado Springs, Colorado, 80913
- PSC
- 6530: HOSPITAL FURNITURE, EQUIPMENT, UTENSILS, AND SUPPLIES
- Last updated
- Oct 29, 2026
- First published
- Last updated on SAM.gov
Identifiers
- Solicitation number
- W81K0026QA200
- Notice ID
- f5b1b23993204a9a99b7ad6adfd7c097
Description
Clinical Salient Requirements:
System must allow clinicians to adjust the focal point and working distance directlyfrom the center sterile handle without moving the light head or leaving the sterilefield.
The light head must utilize vertically segmented reflectors to capture upward-directed light and redirect it downward, creating a crisp, homogenous spot thatminimizes shadows.
The system must offer dual selectable color temperatures (specifically 4100K softwhite and 4500K bright white) adjustable for specific procedure types.
Technological Salient Requirements:
Server Requirements:
The lights must deliver a maximum intensity of 160,000 lux and offer light headdiameter options of 24 inches or 30 inches to fill the surgical cavity with seamlesslight.
The lighting architecture must integrate a manual focus mode of high-intensity LEDscombined with a centrally positioned ring of fixed-focus LEDs.
The lights must be explicitly compatible with and ready for integration with flat-screen displays, booms, and HD surgical cameras.
System Requirement:
The suspension and arm system must provide at least 110 degrees of vertical travelto allow steep downward positioning.
All axes of the lighting fixture must feature 360-degree continuous rotation withouthard spots.
The optical design must allow the fixtures to be placed further away from the patientwhile maintaining precise detail.
Award Overview
Contacts
Contracting Officer
Primary Contact
Documents
Lifecycle
- 1
- 2
- 4Version 4SolicitationArchivedCurrent
L&D Surgical Lights
Similar Opportunities
Additional Details
Solicitation Number
W81K0026QA200
Archive Date
Jul 7, 2026
Office Address
JBSA FT SAM HOUSTON, TX, 78234-4504